SAT 25-JAN-2003 (025) - E-MAIL FROM TERRY -------------------------------------------------------------------
X-Sender: thoulihan@mlml.calstate.edu
X-Mailer: QUALCOMM Windows Eudora Pro Version 4.2.0.58
Date: Wed, 12 Feb 2003 12:53:24 -1000
To: Stephanie Flora
From: Terrence Houlihan
Subject: met station mmob03 (I think?)
Stephanie,
Here is something for your log:
The 3rd deployment of the Coastal Environmental weather station mounted on the MOBY mooring
buoy took place at 0921 23 January 2003 using the R/V Kaimikai-O-Kanaloa. All deployment
operations proceeded smoothly with the exception of a slightly damaged cell phone antenna
which was later replaced. The station began collecting data immediately with all sensors
operating as expected except for a troublesome WET Labs Wetstar fluorometer.
The system is equipped with two R.M.Young model 5103 anemometers, s/n 44265 and 44266, two
Vaisala Humitter 50YX temperature/humidity sensors s/n ?,?, a Seabird Electronics Microcat
IM37 temperature/conductivity sensor s/n 1503, one non functioning WET Labs Wetstar fluorometer
s/n WS1-054, one WET Labs C-Star transmissometer s/n CST-605PB, one Honeywell PPT barometer,
one Navico fluxgate compass and one Coastal Environmental Zeno-3200 data logger and software s/n V1.961_2268-1.0.
The correct wind direction is from sensor #2 and the data is being logged by GMT. The remainder of
the data from the previous deployment is now on the ftp space.
Let me know if you need anything else.
Terry
E-MAIL FROM TERRY -------------------------------------------------------------------
Date: Thu, 13 Feb 2003 10:40:57 -1000
To: Stephanie Flora
From: Terrence Houlihan
Subject: fluorometer
Cc: Dennis.K.Clark@noaa.gov
Stephanie,
I may have found the problem with the fluorometer, it was a formatting problem on my part,
it was only transmitting via cell phone one decimal. I changed the program to transmit
three decimal places, well see how it goes. I also noticed that there seems to be a
pretty steady degradation in the transmissometer signal, must be getting fouled.
Terry
E-MAIL FROM TERRY -------------------------------------------------------------------
X-Sender: thoulihan@mlml.calstate.edu
X-Mailer: QUALCOMM Windows Eudora Pro Version 4.2.0.58
Date: Thu, 13 Feb 2003 12:06:38 -1000
To: Stephanie Flora
From: Terrence Houlihan
Subject: more fluormeter
Stephanie,
The fluorometer data you have for this deployment, the past half day, is no good. This time it
is a calibration problem. I will try to change data logger to the correct values so that it will
output voltage this afternoon when the modem turns on. There is a ten minute window so it is
sometimes hard to catch.
Ill let you know when it is outputting voltage. The transmissometer is outputting voltage and Ill
work on getting you a conversion equation for both.
Terry
PHONE CALL FROM TERRY -------------------------------------------------------------------
14 Feb ~11 am Terry called and said he changed the Fluorometer to output voltage
WED 19-FEB-2003 (050) - A Driver on the MOBY SGI was corrupted and
therefore did not allow MOBY to dial-in. Ray fixed the
problem by rebuilding the driver on Sunday. MOBY dialed in successfully
on Feb 23 the 23 hour file.
FRI 07-MAR-2003 (066) - E-MAIL FROM TERRY -------------------------------------------------------------------
X-Sender: thoulihan@mlml.calstate.edu
X-Mailer: QUALCOMM Windows Eudora Pro Version 4.2.0.58
Date: Fri, 07 Mar 2003 08:49:47 -1000
To: Stephanie Flora
From: Terrence Houlihan
Subject: buoy stuff
Hey Steph,
We just got back from Maui and I did some cleaning on the optical instruments while we
were there, mainly the transmissometer. It was about 2300 on 3/5 gmt. It looks like
it did change something but I will have to wait and see when we get more data.
Terry
TUE 11-MAR-2003 (070) - Sunday 9 Mar the 23 hour file was not sent, on Tuesday I tried
to process the 10 Mar data and neither of the files were readable and the 23 hour
files size was 257281 instead of 255616. The Error message I got from READMOB_ was
??? In an assignment A(matrix,matrix) = B, the number of rows in B
and the number of elements in the A row index matrix must be the same.
Error in ==> C:\matlabR12\mldbase\instrument\MOBY\Readmob_.m
On line 366 ==> Xdat(1:fix(Blue_len/2),Vnum) = fread(fid,[fix(Blue_len/2),1],ushort)
When I went into readmob_ to find the problem the fix(Blue_len/2) was 131072 not 512.
No solution.
FRI 14-MAR-2003 (073) - 23 Hour file, Partial file and unreadable.
READMOB_(03031423.mob,0,mby22201)
! WARNING (Blk1) 14946 Bytes NOT Read!
TUE 25-MAR-2003 (084) - 23 Hour file, Partial file and unreadable.
READMOB_(03032523.mob,0,mby22201)
! WARNING (Blk1) 970 Bytes NOT Read!
SUN 30-MAR-2003 (089) - 20 hour file, Could be a good day, hard to tell.
SAT 12-APR-2003 (102) - Need to start getting Lats and Long from FORTH LOG MAT-file so this needs to exist to process this deployment.
03041623.mob have no Lat, Long or delta time is approximate and the lat and long are the watch circle center.
THU 17-APR-2003 (107) - The MOBY222 was towed to the lee of lanai and recovered (?The next morning?).
E-MAIL FROM MARK -------------------------------------------------------------------
Date: Sat, 19 Apr 2003 08:07:09 -1000
From: Mark Yarbrough
To: Stephanie Flora
Subject: Re: MOBY
Aloha Steph,
We really need to know if the buoy is trying to call into the lab. I am unable
to call into the buoy. This has happened in the past and can last for many
days. In the past however, the buoy was calling out even though I was unable to
call in.
So, can Ray enable the diagnostics so we will know if/when the buoy calls?
As for the last buoy M222, The 22 hour MOBY acquisition was overrunning the
start of the 23 hour file for some reason. Dont know why yet. We will have no
23 hour files for the days you recieved 22 hour files.
Mark
| Sensor | MUX Position | Depth (m) |
|---|---|---|
| Es SFC | 8 | N/A |
| Ed TOP | 10 | 1.212 |
| Lu TOP | 9 | 1.326 |
| Ed MID | 7 | 5.022 |
| Lu MID | 5 | 5.136 |
| Ed BOT | 3 | 8.902 |
| Lu BOT | 4 | 9.016 |
| Eu MOS | 6 | 11.564 |
| LuMOS | 11 | 11.092 |
| BLUE LED | 12 | N/A |
| RED LED | 13 | N/A |
| LAMP | 15 | N/A |
| Name | Code | Conversion |
|---|---|---|
| Unknown | -1 | X; |
| None | 0 | X; |
| Water Temperature (deg C) | 10 | (-4.6789E-16)*X^4 - (1.7279E-11)*X^3 + (7.2224E-07)*X^2 - (1.3381E-02)*X + 9.7741E+01; |
| Blue array temp (deg C) | 11 | (3.0518E-2)*X; |
| Red array temp (deg C) | 12 | (3.0518E-2)*X; |
| Calib src block temp (deg C) | 13 | (-4.6789E-16)*X^4 - (1.7279E-11)*X^3 + (7.2224E-07)*X^2 - (1.3381E-02)*X + 9.7741E+01; |
| Depth Sensor temp (deg C) | 14 | (-4.6789E-16)*X^4 - (1.7279E-11)*X^3 + (7.2224E-07)*X^2 - (1.3381E-02)*X + 9.7741E+01; |
| Blue CCD head temp (deg C) | 15 | (-4.6789E-16)*X^4 - (1.7279E-11)*X^3 + (7.2224E-07)*X^2 - (1.3381E-02)*X + 9.7741E+01; |
| Red CCD head temp (deg C) | 16 | (-4.6789E-16)*X^4 - (1.7279E-11)*X^3 + (7.2224E-07)*X^2 - (1.3381E-02)*X + 9.7741E+01; |
| TT7 board temp (deg C) | 17 | (-1.1978E-13)*X^4 - (1.1059E-09)*X^3 + (1.1556E-05)*X^2 - (5.3524E-02)*X + 9.7741E+01; |
| TT7 case temp (deg C) | 18 | (-1.1978E-13)*X^4 - (1.1059E-09)*X^3 + (1.1556E-05)*X^2 - (5.3524E-02)*X + 9.7741E+01; |
| Depth (meters) | 20 | (5.8710E-09)*X^2 + (1.2780E-02)*X - 4.3436E-03; |
| X Tilt (degrees) | 21 | (-2.8357E-13)*X^3 + (1.2891E-10)*X^2 + (3.0807E-03)*X + 3.4178E-02; |
| Y Tilt (degrees) | 22 | (1.6168E-13)*X^3 + (6.7126E-11)*X^2 + (3.0191E-03)*X + 4.4186E-03; |
| Inclination | 2122 | X; |
| Blue cal diode mon (volts) | 23 | (3.0518E-4)*X; |
| Red cal diode mon (volts) | 24 | (3.0518E-4)*X; |
| Blue heat mon (volts) | 25 | (3.0518E-4)*X; |
| Red heat mon (volts) | 26 | (3.0518E-4)*X; |
| Coolant flow (liters/min) | 27 | (6.1035E-5)*X; |
| System current (amps) | 28 | (3.05E-3)*X - 40; |
| TT7 Mains (volts) | 29 | (6.105E-3)*X; |
| TT7 case humidity (%RH) | 30 | (X*0.02906-19)/(1.054-.00216*T); |
| Name | Code | Conversion |
|---|---|---|
| TT7 upper bat (volts) | 100 | 6.104E-03*X |
| TT7 board temp (deg C) | 101 | -1.1978E-13*X^4 -1.1059E-09*X^3 +1.1556E-05*X^2 -5.3524E-02*X +9.7741E+01; |
| TT7 case temp (deg C) | 102 | -1.1978E-13*X^4 -1.1059E-09*X^3 +1.1556E-05*X^2 -5.3524E-02*X +9.7741E+01; |
| TT7 case humidity (%RH) | 103 | (X*0.02906-19)/(1.054-.00216*T); |
| Solar panel #1 current (Amps) | 104 | 3.052E-4*X; |
| Solar panel #2 current (Amps) | 105 | 3.052E-4*X; |
| Solar panel #3 current (Amps) | 106 | 3.052E-4*X; |
| Solar panel #4 current (Amps) | 107 | 3.052E-4*X; |
| Upper bat. charge current (Amps) | 108 | 6.104-4*X; |
| Upper arm collector depth (Meters) | 109 | (0.0003*(X-1.570) - 0.9435); % 1.088 is the zero volts |
| Upper bat. drain current (Amps) | 110 | 6.104-4*X; |
| N/A | 111 | (NaN); |
| TT5 mains power (Volts) | 112 | 6.104E-03*X; |
| Lower bat. charge current (Amps) | 113 | 3.662E-3*X; |
| MOS current (Amps) | 114 | 3.662E-3*X; |
| MOS voltage (Volts) | 115 | 6.104E-03*X; |
| Lower Battery #1 (Volts) | 116 | 1.221E-2*X; |
| Lower Battery #2 (Volts) | 117 | 1.221E-2*X; |
| Lower Battery #3 (Volts) | 118 | 1.221E-2*X; |
| Lower Battery #4 (Volts) | 119 | 1.221E-2*X; |


File: 222_RSP.MLD
DscS 14: MOBY222 Nominal Integration Time (sec)
DscS 15: MOBY222 Calibrated Integration Time (sec)
DscS 16: MOBY222 Nominal CCD Bin Factor (pix)
DscS 17: MOBY222 Calibrated CCD Bin Factor (pix)
Xdat:
NomInt, CalInt, NomBin, CalBin
0.050000, 0.069651, 1.000000, 1.000000
0.100000, 0.118030, 2.000000, 2.000000
0.200000, 0.214983, 3.000000, 3.000000
0.250000, 0.263570, 4.000000, 4.000000
0.300000, 0.312232, 6.000000, 6.000000
0.400000, 0.409755, 8.000000, 8.000000
0.500000, 0.507498, 12.000000, 12.000000
0.600000, 0.605415, 16.000000, 16.000000
0.700000, 0.703469, 24.000000, 24.000000
0.800000, 0.801632, 32.000000, 32.000000
0.900000, 0.899883, 48.000000, 48.000000
1.000000, 0.998204, 64.000000, 64.000000
1.600000, 1.589047, 96.000000, 96.000000
2.000000, 1.983385,128.000000,128.000000
3.000000, 2.969611,192.000000,192.000000
3.200000, 3.166858,384.000000,384.000000
4.000000, 3.955741, ,
5.000000, 4.941545, ,
6.000000, 5.927016, ,
6.400000, 6.321121, ,
7.000000, 6.912202, ,
8.000000, 7.897168, ,
9.000000, 8.881975, ,
10.000000, 9.866680, ,
12.799999, 12.623697, ,
16.000000, 15.775045, ,
20.000000, 19.715988, ,
25.599998, 25.237938, ,
30.000000, 29.580973, ,
32.000000, 31.556406, ,
40.000000, 39.466446, ,
50.000000, 49.372002, ,
51.199997, 50.561935, ,
60.000000, 59.295845, ,
64.000000, 63.270035, ,
70.000000, 69.235870, ,
80.000000, 79.189987, ,
90.000000, 89.156258, ,
102.399994,101.528618, ,
128.000000,127.109253, ,
256.000000,255.201874, ,
512.000000,508.916229, ,
-------------------------------------------------------------------------------------------------- MOBY222: Table of Engineering Information -------------------------------------------------------------------------------------------------- HOUR: 20 | | LITE DARK | | Int Bin | Maximum Maximum | No. No. | Time Factor | Counts Counts | Lite Dark Date Source | (B/R) (B/R) | (B/R) (B/R) | -------------------------------------------------------------------------------------------------- 2003106 LuTop | 2.00/40.00 128/384 | 45068/26390 1214/6304 | 5 2 2003106 LuMid | 2.00/64.00 128/384 | 37848/18288 1237/9978 | 5 2 2003106 LuBot | 2.00/64.00 128/384 | 42239/16190 1216/9788 | 5 2 2003106 LuMOS | 0.80/10.00 8/384 | 38733/46632 983/2400 | 5 2 2003106 EdTop | 0.75/0.60 32/64 | 32548/26432 1007/1040 | 5 2 2003106 EdMid | 0.75/2.50 32/64 | 31844/20066 1005/1106 | 5 2 2003106 EdBot | 0.75/2.00 32/384 | 28802/25692 1002/1346 | 5 2 2003106 EuMOS | 8.00/32.00 128/384 | 53137/6024 1927/5518 | 3 2 2003106 EsSfc | 0.75/0.75 16/16 | 32505/24637 989/1020 | 3 2 2003106 Blue LED | 1.00/1.00 64/384 | 30967/1940 1061/1218 | 3 2 2003106 Red LED | 1.00/1.00 384/64 | 15055/31294 1437/1069 | 3 2 2003106 Lamp | 1.00/1.00 64/16 | 27783/40118 1054/1053 | 3 2 -------------------------------------------------------------------------------------------------- HOUR: 22 | | LITE DARK | | Int Bin | Maximum Maximum | No. No. | Time Factor | Counts Counts | Lite Dark Date Source | (B/R) (B/R) | (B/R) (B/R) | -------------------------------------------------------------------------------------------------- 2003106 LuTop | 3.50/64.00 64/384 | 43101/60516 1191/9834 | 6 2 2003106 LuMid | 3.50/64.00 64/384 | 43047/24166 1217/10468 | 6 2 2003106 LuBot | 3.50/64.00 64/384 | 37023/15982 1194/10158 | 6 2 2003106 LuMOS | 0.45/1.10 16/384 | 49847/9944 1016/1246 | 6 2 2003106 EuMOS | 8.00/32.00 128/384 | 56310/6180 2007/5664 | 3 2 2003106 EsSfc | 0.75/0.75 16/16 | 42139/31630 991/1020 | 3 2 2003106 Blue LED | 1.00/1.00 64/384 | 30876/1914 1070/1224 | 3 2 2003106 Red LED | 1.00/1.00 384/64 | 14948/31076 1479/1077 | 3 2 2003106 Lamp | 1.00/1.00 64/16 | 27691/40099 1063/1030 | 3 2 -------------------------------------------------------------------------------------------------- HOUR: 23 | | LITE DARK | | Int Bin | Maximum Maximum | No. No. | Time Factor | Counts Counts | Lite Dark Date Source | (B/R) (B/R) | (B/R) (B/R) | -------------------------------------------------------------------------------------------------- 2003107 LuTop | 2.00/40.00 128/384 | 45773/50960 1245/6668 | 5 2 2003107 LuMid | 2.00/64.00 128/384 | 48085/20684 1298/11076 | 5 2 2003107 LuBot | 2.00/64.00 128/384 | 41198/15974 1251/10044 | 5 2 2003107 LuMOS | 0.80/10.00 8/384 | 41485/40824 983/2458 | 5 2 2003107 EdTop | 0.75/0.60 32/64 | 28149/21395 1008/1035 | 5 2 2003107 EdMid | 0.75/2.50 32/64 | 34254/19566 1005/1100 | 5 2 2003107 EdBot | 0.75/2.00 32/384 | 26206/20156 1006/1478 | 5 2 2003107 EuMOS | 8.00/32.00 128/384 | 48439/6112 2003/5618 | 3 2 2003107 EsSfc | 0.75/0.75 16/16 | 40584/30841 993/1020 | 3 2 2003107 Blue LED | 1.00/1.00 64/384 | 30785/1924 1060/1224 | 3 2 2003107 Red LED | 1.00/1.00 384/64 | 15010/31051 1493/1055 | 3 2 2003107 Lamp | 1.00/1.00 64/16 | 27821/40193 1056/1054 | 3 2 --------------------------------------------------------------------------------------------------